Comparison of Digital PCR and Quantitative PCR with Various SARS-CoV-2 Primer-Probe Sets
1Microbiological Analysis Team, Biometrology Group, Korea Research Institute of Standards and Science (KRISS), Daejeon 34113, Republic of Korea
2National Research Laboratory of Molecular Microbiology and Toxicology, Department of Agricultural Biotechnology, Seoul National University, Seoul 08826, Republic of Korea
3Center for Convergent Research of Emerging Virus Infection, Korea Research Institute of Chemical Technology, Daejeon 34114, Republic of Korea
4College of Pharmacy, Chungnam National University, Daejeon 34134, Republic of Korea
5Department of Bio-Analytical Science, University of Science and Technology (UST), Daejeon 34113, Republic of Korea
6Research Center for Bioconvergence Analysis, Korea Basic Science Institute, Cheongju 28119, Republic of Korea
7Research Group of Food Processing, Korea Food Research Institute, Wanju-gun, Jeollabuk-do 55365, Republic of Korea
8Department of Predictive Toxicology, Korea Institute of Toxicology, Daejeon 34114, Republic of Korea
9Department of Biological Science, Chungnam National University College of Bioscience and Biotechnology, Daejeon 34134, Republic of Korea
10Zoonotic Virus Laboratory, Institut Pasteur Korea, Seongnam-si, Gyeonggi-do 13488, Republic of Korea
11Department of Internal Medicine, Jeonbuk National University Medical School, Jeonju 54986, Republic of Korea
12Biomedical Research Institute of Jeonbuk National University Hospital, Jeonju 54907, Republic of Korea
J. Microbiol. Biotechnol. 2021; 31(3): 358-367
Published March 28, 2021 https://doi.org/10.4014/jmb.2009.09006
Copyright © The Korean Society for Microbiology and Biotechnology.
Abstract
Keywords
Graphical Abstract
Introduction
Coronaviruses have caused several notable respiratory disease outbreaks in humans, such as severe acute respiratory syndrome (SARS). The SARS coronavirus was the first fatal pathogenic human coronavirus to be identified and caused 8,096 infections and 774 deaths in 26 countries in 2003 [1]. Prior to the emergence of SARS-CoV, only two HCoVs (HCoV-229E and HCoV-OC43) were known, both of which caused mild respiratory symptoms [2]. Bats were the natural hosts of SARS-CoV, and civet cats were regarded as the intermediate host [3]. Extensive studies on bat coronaviruses showed that such coronaviruses could be potential human pathogens [4, 5]. MERS-CoV was first reported in 2012 and caused 2,494 infections and 858 deaths in 27 countries as of November 2019 (WHO, Middle East respiratory syndrome coronavirus (MERS-CoV)). MERS-CoV also originated from bat coronaviruses, and camels were the intermediate host [7].
According to the World Health Organization (WHO), the WHO China Country Office was informed of cases of pneumonia of unknown etiology in Wuhan City, Hubei Province, on December 31st, 2019 [8]. The novel coronavirus, currently termed SARS-CoV-2, was officially announced as the causative agent by the International Committee on Taxonomy of Viruses (ICTV) [9]. SARS-CoV-2, or severe acute respiratory syndrome coronavirus 2, has resulted in over fifty million laboratory-confirmed cases and over a million deaths as of November 2020 [10]. A viral genome sequence was released for immediate public health support via the community online resource virological.org on January 10th (Wuhan-Hu-1, GenBank accession number MN908947) [3], followed by four other genomes that were deposited on January 12th in the viral sequence database curated by the Global Initiative on Sharing All Influenza Data (GISAID). The genome sequences suggest the virus is closely related to the members of a viral species termed severe acute respiratory syndrome (SARS)-related CoV, a species defined by the agent of the 2002/03 outbreak of SARS in humans [4, 5]. The species also comprises of a large number of viruses mostly detected in Rhinolophid bats in Asia and Europe. For the past several decades, quantitative polymerase chain reactions (qPCR) have become the gold standard for quantifying relative gene expressions. On the other hand, the recently developed digital polymerase chain reaction (dPCR) enables a highly sensitive measurement and absolute quantitation of nucleic acids [11, 12]. Droplet dPCR (ddPCR) can provide absolute quantification of target DNA or RNA molecules without external reference materials. The PCR reaction mixture with target nucleic acid molecules are portioned into thousands of nanoliter-sized water-in-oil emulsion droplets. These droplets are amplified with PCR, and each individual droplet is analyzed. The droplets with a template are counted as PCR-positive droplets, and the droplets without template are PCR-negative droplets. The absolute quantification of target nucleic acids can be done based on the number of positive and negative droplets. ddPCR has an advantage for the clinical diagnosis of SARS-CoV-2 to reduce the false negative results [13], which could be highly sensitive compared to real time PCR (RT-PCR) [14-17]. Moreover, there are some research groups that have analyzed many samples from COVID-19 confirmed patients by droplet digital PCR (ddPCR) and RT-PCR based on two target genes (ORF1ab and N). The overall concept looks similar to analyzing a large number of COVID-19 patient samples using digital PCR [18, 19]. However, none of these groups applied the ddPCR reference value using the reference material in the patient sample to measure an accurate quantification and to achieve diagnostic reliability [13,20-24]. DNA or RNA quantities should be metrologically traceable to a reference [25, 26]. As of this writing, multiple nucleic acid quantitation methods have been developed, such as chemical analysis methods based on isotope-dilution mass spectrometry (IDMS), capillary electrophoresis (CE), and enumeration-based flow cytometric method (FCM) counting [27, 28]. These methods can be accurately calibrated with solutions of nucleic acids. In addition, an international comparison study was performed between national metrology institutes (NMI) using the ddPCR method [29]. Recently, the droplet digital PCR (ddPCR) method has emerged as a powerful analytical technique for clinical utility [30, 31]. For example, ddPCR can be used for the rapid enumeration of viral genomes and particles [32-34].
In the present case of SARS-CoV-2, virus isolates or samples from infected patients have yet to be made available to the international public health community. In this study, we concluded that the optimized conditions are required to increase the precision of ddPCR to develop reference materials with matrix conditions.
Materials and Methods
Construction of RNA Standards
Full-length N and E genes of SARS-CoV-2 (GeneBank: MN908947) were synthesized and cloned in the pET21a plasmid. The plasmids were amplified with the T7 promoter primer (5’-TAATACGACTCACTATAGGG-3’) and T7 terminator primer (5’-GCTAGTTATTGCTCAGCGG-3’). The amplicons were spectrophotometrically quantified at 260 nm. A total of 200 ng of the PCR product was used for the in vitro transcription (MEGAscript T7 Transcription kit; Thermo Fisher Scientific, USA), which was performed at 37°C overnight in a 20 μl reaction mixture containing 2 μl of reaction buffer, 2 μl of each nucleoside triphosphate, and 2 μl of enzyme mix. The template DNAs were removed by digestion with 2 U of Turbo DNase I for 15 min at 37°C. The RNAs were precipitated by adding 6 μl of 3 M sodium acetate and 150 μl of 98% ethanol, which was followed by a subsequent incubation at -20°C for 30 min. After 15 min of centrifugation at 17,000 rpm, the supernatant was removed and 200 μl of 70% ethanol was added. After another 10-min centrifugation at 17,000 rpm, the supernatant was removed, and the pellet was dissolved in 20 μl of RNase-free H2O (Takara Bio Inc., Japan). Quantitation of the RNAs was performed spectrophotometrically at 260 nm. The measurement of the RNA concentrations was performed in duplicate, and the concentration was then converted to the molecule number [35].
Cell and Virus RNA Extraction
The human cell lines were obtained from the relevant culture collections Vero E6 cells (ATCC CRL-1586) and MRC-5 cells (ATCC CCL-171) were obtained from the American Type Culture Collection (ATCC), and Huh-7 cells (JCRB 0403) were obtained from the Japan Cell Research Bank (JCRB). The cells were cultured at 37°C in a 5% CO2 incubator unless described otherwise.
The clinical isolate SARS-CoV-2 (NCCP 43326) was obtained from the National Culture Collection for Pathogens (NCCP). The strains were introduced into the Vero E6 cells (ATCC CRL-1586) and maintained in Dulbecco’s modified Eagle medium (DMEM) supplemented with 2% (v/v) fetal bovine serum (FBS) and 1%antibiotic-antimycotic solution (Thermo Fisher Scientific). The SARS coronavirus HKU-39849 (GenBank:AY278491.2, provided by Dr. Malik Peiris of the University of Hong Kong) was inoculated into the Vero E6 cells. Cytopathic effects were observed two days after inoculation and the viral titer was determined via a plaque assay. The patient-derived isolate MERS-CoV strain KNIH/002_05_2015 [36] was obtained from the KNIH and inoculated into cultured Huh-7 cells. The Huh-7 cells were maintained in Dulbecco’s modified Eagle’s medium (DMEM, HyClone, USA) supplemented with 10% fetal bovine serum (FBS, HyClone, USA) at 37°C in a 5%CO2 incubator. HCoV OC43 (ATCC VR1588) was obtained from the ATCC and inoculated into the MRC-5 cells. The culture media of the MRC-5 cells were Gibco Minimum Essential Media (MEM; Thermo Fisher Scientific) supplemented with 10% heat-inactivated Fetal Bovine Serum (FBS), 1% Sodium Pyruvate, 1% penicillin, and 1%non-essential amino acids. The inoculated cells were cultured with virus growth medium (MEM with 2% heat-inactivated FBS, 1% penicillin, and 1% non-essential amino acids) at 33°C. The supernatant of the culture medium was stored at -80°C until RNA extraction. The viral RNAs of SARS-CoV-2, MERS-CoV and HCoV-OC43 were extracted using the QIAamp Viral RNA Mini Kit (Qiagen, Germany) according to the manufacturer’s instructions. The viral RNA of SARS-CoV was extracted from the culture medium using the MagMAX-96 viral RNA isolation kit (Thermo Fisher Scientific) according to the manufacturer's instructions.
Clinical Samples and RNA Preparation
The clinical samples used in this study were collected from subjects according to registered protocols approved by the Institutional Review Board (IRB) of Jeonbuk National University Hospital with all patients having signed written informed consent forms (IRB registration number: CUH 2020-02-050-008). The clinical characteristics of the patients are shown in Table S1. Upper respiratory tract specimens (naso- and oropharyngeal swabs) from COVID-19 patients were suspended in a transport medium (eNAT; Copan, USA) and stored at -80°C until use. The RNA extraction from the clinical samples was performed using the QIAamp Viral RNA Mini Kit (Qiagen) according to the manufacturer’s instructions. The extracted viral RNA was used as a template for qPCR assays and cDNA synthesis. The qPCR assays with undiluted RNA were done using the Allplex SARS-CoV-2 Assay kit (Seegene, Korea) according to the manufacturer’s instruction. The Ct values from extracted RNA were shown in Table S1.
Primer and Probes
The primer-probe sets used in this study are listed in Table 1. For the detection of SARS-CoV-2, primer-probe sets of from the Centers for Disease Control and Prevention (CDC, USA), the University of Hong Kong (HKU, Hong Kong), the National Institute of Infectious Disease Department of Virology III (NIID, Japan), and the National Institute of Health (NIH, Thailand) were selected according to laboratory guidance from WHO [37]. The specific primer-probe sets for MERS-CoV and HCov-OC43 were used for the quantification of the viral RNA genomes [38, 39]. All primers and probes were synthesized by Neoprobe (Korea). All probes were labeled with the reporter molecule 6-carboxyfluorescein (FAM) at the 5'-end and with the quencher Black Hole Quencher 1 (BHQ-1) at the 3'-end.
-
Table 1 . Information of primers and probes used in the study.
Institute Gene Name Type* Sequence (5’ → 3’) Taxon Ref. CDC (USA) N 2019-nCoV_N1 F GACCCCAAAATCAGCGAAAT SARS-CoV-2 [37] R** TCTGGTTACTGCCAGTTGAATCTG P+ ACCCCGCATTACGTTTGGTGGACC 2019-nCoV_N2 F TTACAAACATTGGCCGCAAA SARS-CoV-2 R GCGCGACATTCCGAAGAA P ACAATTTGCCCCCAGCGCTTCAG NIID (Japan) NIID_2019-nCoV_N F AAATTTTGGGGACCAGGAAC SARS-CoV-2 R TGGCAGCTGTGTAGGTCAAC P ATGTCGCGCATTGGCATGGA NIH (Thailand) WH-NIC N-R F CGTTTGGTGGACCCTCAGAT SARS-CoV-2 R CCCCACTGCGTTCTCCATT P CAACTGGCAGTAACCA HKU (Hong Kong) HKU-N F TAATCAGACAAGGAACTGATTA Sarbeco R CGAAGGTGTGACTTCCATG P GCAAATTGTGCAATTTGCGG Charité (Germany) E E_Sarbeco F ACAGGTACGTTAATAGTTAATAGCGT Sarbeco [41] R ATATTGCAGCAGTACGCACACA P ACACTAGCCATCCTTACTGCGCTTCG University of Bonn Medical Centre (Germany) upE F GCAACGCGCGATTCAGTT MERS-CoV [38] R GCCTCTACACGGGACCCATA P CTCTTCACATAATCGCCCCGAGCTCG University of Leuven (Belgium) OC43 F ATGTTAGGCCGATAATTGAGGACTAT HCoV-OC43 [39] R AATGTAAAGATGGCCGCGTATT P CATACTCTGACGGTCACAAT *F: forward primer, R: reverse primer, P: Probe
cDNA Synthesis of Viral RNA Genomes
The cDNAs of the clinical and cultured (SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43) viral RNA genomes were synthesized using the LunaScript RT SuperMix Kit (New England Biolabs, USA). The reaction mixture had a volume of 20μl and consisted of 4μl of 5X LunaScript RT Supermix, 1μl of RNA template, and 15 μl of distilled water. The reactions were carried out according to the manufacturer’s instructions. The synthesized cDNAs were serially diluted with distilled water and used as templates for qPCR and ddPCR without purification.
qPCR and Droplet Digital PCR Measurement
The templates were serially diluted to a few copy numbers/μl. To remove the bias from different reverse transcription steps, the same cDNAs from the viral RNA were used as templates for both qPCR and ddPCR. RT-qPCR analysis was performed on the StepOne and StepOnePlus Real-Time PCR system (Thermo Fisher Scientific). The reaction mixture had a volume of 25μl and consisted of 12.5μl of maxima Probe/ROX qPCR master mix (2X) (Thermo Fisher Scientific), 1μl of cDNA template, 1μl of 10μM forward primer, 1μl of 10μM reverse primer, 1μl of 5 μM probe labeled with FAM, and 8.5 μl of distilled water. The ddPCR analysis was performed using a QX200 system (BioRad Laboratories, USA). The reaction mixture had a volume of 20μl and consisted of 10μl of ddPCR supermix for probes (BioRad Laboratories), 1μl of cDNA template, 1 μl of 10μM forward primer, 1μl of 10μM reverse primer, 1μl of 5 μM probe labeled with FAM, and 6 μl of distilled water. All valid copy numbers were selected according to the manufacturer's instructions [40].
Data Analysis
The qPCR data were initially analyzed using the StepOne software (Thermo Fisher Scientific). Raw data (
Results
Quantification of Viral and IVT RNAs
In this study, the primer-probe sets of each E gene were used to quantify the viral genomic RNA of SARS-CoV-2 and related coronaviruses in triplicate (Fig. 1 and Table 1) [37-39,41]. SARS-CoV, MERS-CoV, and HCoV-OC43 were used a negative control for the assessment of SARS-CoV-2 specific primer-probe sets. The copy numbers of SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 viral RNA were measured using ddPCR, and the measured values of the viral genomic RNA were 8.04 × 105, 6.52 × 105, 9.37 × 104, and 9.57 × 104 copies/μl, respectively. The standard curves of qPCR and ddPCR are linear, indicating the amplification efficiency of both methods is consistent (Fig. 1). The quantified viral genomic RNAs of SARS-CoV, MERS-CoV, and HCoV-OC43 were used as negative templates for SARS-CoV-2 specific assays. To assess the quantification of the viral genomic RNA and partial synthetic viral RNA, in vitro transcript (IVT) N and E gene RNA were measured via qPCR and ddPCR in triplicate (Fig. 2). The copy numbers of N IVT RNA and E IVT RNA were measured as 1.8 × 109 and 3.47 × 109, respectively. The amplification efficiencies were calculated from the standard curves of SARS-CoV-2 viral RNA and IVT RNAs. In the qPCR reactions, each E gene primer-probe sets targeting the SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 showed Ct values of 28.64, 27.96, 26.96, and 27.69, respectively. In addition, the ddPCR reactions with each E gene primer-probe set targeting SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 showed copy numbers of 2.68 × 106, 4.69 × 106, 7.85 × 105, and 2.66 × 105, respectively.
-
Fig. 1.
Detection of four viruses based on E gene assays by qPCR and ddPCR. (A ) and (B ): The Ct value and copy number of SARS-CoV-2 and SARS-CoV using the E sarbecovirus assay. (C ) and (D ): The Ct value and copy number of MERS and OC43 virus using the upE and OC43_E assay, respectively. All template were performed in triplicate and points indicates average value.
-
Fig. 2.
Detection of the IVT gene by qPCR and ddPCR. (A ): The Ct value and copy number of the IVT N gene using the 2019-nCoV_N2 assay. (B ): The Ct value and copy number of the IVT E gene using the E sarbecovirus assay. All template were performed in triplicate and points indicates average value.
Primer-Probes Validation
To assess the specificity and sensitivity of the five SARS-CoV-2 specific N gene targeting primer-probe sets (2019-nCoV_N1, 2019-nCoV_N2, NIID_2019-nCoV_N, HKU-N, WH-NIC N, the assays were done with SARS-CoV-2 RNA and other human pathogenic corona viral RNAs (SARS-CoV, MERS-CoV, and HCoV-OC43). The qPCR and ddPCR reactions using the N gene targeting primer-probes sets were positive in the SARS-CoV-2 genomic RNA in triplicate (Fig. 3 and Table 3). The reactions were negative with the other betacorona viral genomic RNAs such as SARS-CoV, MERS-CoV, and HCoV-OC43. The reaction with the E primer-probe sets was positive for each target viral RNA. The reactions using
-
Table 3 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays.
Dilution factor 2019-nCoV_N1 2019-nCoV_N2 NIID_2019-nCoV_N WH-NIC N HKU-N qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR 10-2 18.89 S* 18.87 S 18.14 S 21.40 S 20.79 S 10-3 22.16 21173 22.19 20340 21.63 15173 24.65 19993 24.31 18473 10-4 25.99 1907 26.06 1659 25.40 1177 28.52 1599 28.09 1553 10-5 30.66 135 30.60 132 29.93 60 33.11 105 32.68 105 10-6 34.12 11 34.57 8 34.91 7 36.71 5 36.71 12 R2 0.9968 0.9973 0.9943 0.9971 0.9978 *Saturated
**All template were performed in triplicate and data in table indicates average value.
-
Fig. 3.
Comparison of the N gene primer-probe sets for the qPCR and ddPCR methods. For the comparison of viral SARS-CoV-2, primer-probe sets from the Centers for Disease Control and Prevention (CDC, USA) (A andB ), the National Institute of Infectious Disease Department of Virology III (NIID, Japan) (C ), the University of Hong Kong (HKU, Hong Kong) (D ), and the National Institute of Health (NIH, Thailand) (E ) were used based on the results of Table 4. All template were performed in triplicate and points indicates average value.
IVT RNAs as a Standard for RT-qPCR
The R2 values of the viral RNA and IVT RNA (N genes) standard curve from the RT-qPCR assays were calculated as 0.9973 and 0.9987, respectively, whereas the R2 values of the viral RNA and IVT RNA (E genes) standard curve from the RT-qPCR assays were calculated as 0.9956 and 0.9993, respectively (Figs. 1 and 2 and Tables 2 and 3). These values indicate that the standard curves of both the viral and IVT RNAs are linear and accurate. The amplification efficiency of both the viral and IVT RNAs were calculated from the slopes of the standard curves. The efficiencies of both the viral and IVT RNAs according to the target genes were similar (Figs. 1 and 2). The similar amplification efficiencies show that the quantified IVT RNA can be used as a standard for RT-qPCR measurement of viral RNA.
-
Table 2 . Validation of the SARS-CoV-2 specific N gene targeting primer-probe sets used in the study.
Template SARS-CoV-2 (15 pg/μl) SARS-CoV (8.7 pg/μl) MERS-CoV (72 pg/μl) HCoV-OC43 (60 pg/μl) Primer-probes qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 2019-nCoV_N1 26.30 6123 Undetermined 0 Undetermined 0 Undetermined 0 2019-nCoV_N2 26.57 5340 Undetermined 2.2 Undetermined 4.3 39.09 4.8 NIID_2019-nCoV_N 26.89 3993 Undetermined 0 Undetermined 0 Undetermined 0 HKU-N 27.57 4863 27.24 6030 Undetermined 0 Undetermined 0.8 WH-NIC N 28.91 6153 Undetermined 0 Undetermined 0 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value.
Comparison of qPCR and ddPCR Assays for COVID-19 Clinical Samples
Both the qPCR and ddPCR assays were tested with five SARS-CoV-2 positive and three SARS-CoV-2 negative clinical samples. The clinical characteristics of the confirmed COVID-19 patients with symptoms are summarized in Table S1. The positive patients were determined in clinics with the qPCR assay. The results from the positive samples showed that both qPCR and ddPCR can detect up to approximately 10 copy/μl of SARS-CoV-2 genomic RNA (Table 5). The control assays without templates did not produce any signals from both the qPCR and ddPCR assays. However, the qPCR and ddPCR assays with three negative patient samples (N1, N3, and N5) showed a high Ct value or small amount of copy number in some reactions (Table 4). According to the qPCR and ddPCR results with templates of very low concentrations, the signals from assays with a template of very low concentration can be difficult to distinguish from background signals. These results indicated that the sensitivity of ddPCR can be equal to or more than that of the optimized qPCR.
-
Table 4 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays on negative patient sample.
Primer-Probes 2019-nCoV_N2 primer-probe set WH-NIC N primer-probe set Template qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) N1 (153ng/μl) Undetermined 2.8 Undetermined 0 N3 (101ng/μl) Undetermined 4.3 Undetermined 0 N5 (243ng/μl) Undetermined 0.5 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value.
-
Table 5 . Results of the Ct values by qPCR and copy number by ddPCR on patient sample.
Primer-probe 2019-nCoV_N2 primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 26.87 1051 35.03 7.2 35.15 2.4 Undetermined 0 35.05 7.53 10-2 30.52 82 37.03 1.33 Undetermined 0.73 Undetermined 0.53 Undetermined 1.2 10-3 33.81 5.93 Undetermined 0.67 Undetermined 0.6 Undetermined 0 Undetermined 1.07 Primer-probe WH-NIC N primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 28.77 836.00 36.11 5 35.88 6.80 Undetermined 1.40 35.86 7.67 10-2 32.25 107.33 38.54 3 38.52 1.93 Undetermined 0.00 37.02 1.87 10-3 35.95 10.40 Undetermined 2.40 38.27 2.07 38.44 1.27 Undetermined 0.47 *The initial concentration of RNA templates were 46.9, 21.1, 27.5, 17.2, and 15.9 ng/μl, respectively.
**All template were performed in triplicate and data in table indicates average value.
+Only one of the triplicate was positive (single droplet)
Discussion
The outbreak of COVID-19 is a pandemic threat caused by the emergence of SARS-CoV-2, a newly discovered human coronavirus [42, 43]. SARS-CoV-2 has four structural proteins, known as the S (spike), E (envelope), M (membrane), and N (nucleocapsid) proteins [44]. The SARS-CoV-2 S protein binds with angiotensin-converting enzyme 2 (ACE2) with high-affinity, which leads to clinical features of pneumonia in many patients [45, 46]. SARS-CoV-2 is highly contagious and has quickly spread across the world, which further emphasizes the essential role of diagnostics in the control of communicable diseases [47].
Currently, a reliable test (also the most widely used) for detecting acute infections of SARS-CoV-2 is the reverse transcription-quantitative polymerase chain reaction (RT-qPCR) method [48]. Multiple national laboratories have developed assays targeting conserved regions that could potentially detect pancoronaviruses [49, 50]. A previous study provided information such as the specific primer and probe sequence, thermal profile, and reagents for the RT-qPCR to optimize the reaction conditions for the detection of SARS-CoV-2 RNA [51]. ddPCR is a method that improves upon conventional PCR methods that can individually amplify and directly quantify pathogen RNA or DNA [52]. Due to the principle of ddPCR, the signal from each droplet with a template can be clearly distinguished from the noise from the droplet without a template. Several studies have compared RT-qPCR and ddPCR for the quantification of viral genomes [53-55].
In this study, SARS-CoV-2 RNA concentrations were measured using the qPCR and ddPCR methods. The qPCR results showed that the Ct value of a given template was greatly dependent on the individual primer-probe sets. However, the copy numbers determined by ddPCR were relatively stable and reliable in the quantification of viral RNA samples, regardless of the primer-probe sets. A previous study reported that qPCR and ddPCR results were consistent for 95 positive samples and that there was a high correlation between the Ct values of qPCR and the copy number values of ddPCR among patient samples [56]. Although the Ct values of each primer-probe set exhibited variance, the copy numbers from ddPCR showed that N genes are more abundant than E genes. These findings are also consistent with another previous study, indicating N genes are good targets for SARS-CoV-2 diagnostics [57].
The qPCR method requires the use of a standard curve to obtain quantitative measurements of an unknown sample [58]. However, in the case of ddPCR, the distribution of a target-specific amplification into partitions is calculated using the Poisson distribution, enabling the absolute quantification of the target gene based on the ratio of positive against all partitions at the end of the reaction without dependence on an external standard [59]. In addition to these advantages, our results revealed that ddPCR is a sensitive and accurate method for SARS-CoV-2 detection in low viral copies (Fig. 3). Further experiments should be performed even if patients are cured of COVID-19, and such tests should be conducted using ddPCR to lower the limit of detection (LOD) to be able to detect low amounts of viral RNA.
The Coronavirus Standards Working Group led by the Joint Initiative for Metrology in Biology (JIMB) is conducting international collaborative experiments to develop guidelines for the purpose of evaluating and establishing the availability of common and appropriate standards, diverse reference materials, validation tests, and reference measurement protocols that can enable reliable COVID-19 testing [60, 61]. Our results serve as a guide for the standardization of analytical methods and the development of reference materials that are required for accurate COVID-19 diagnosis based on both RT-qPCR and ddPCR. Although, ddPCR-based diagnostics can serve as a companion method to the current standard RT-qPCR to provide sensitive and accurate quantitative results for emergency testing that can be adopted for clinical use, there are some drawbacks. The narrow dynamic range and high capital and operational cost of ddPCR are the reasons preventing the deployment of ddPCR in clinical fields. Due to the nature of ddPCR, the precise quantification can be done with a decent number of positive droplets. ddPCR can be a precise diagnostics method, but there are some drawback. Therefore, the combination of both qPCR and ddPCR could be better methods for the SARS-CoV-2 diagnostics; qPCR assay with ddPCR defined reference materials could be precise and practical assays for pathogen detection.
Conclusion
Our study demonstrated that the Ct values of samples from qPCR assays were significantly variable depending on the sequences of primer and probe sets, whereas the copy numbers of the samples from the ddPCR assays were hardly affected by the sequences of the primer and probe sets. These results indicate that ddPCR can be performed with sub-optimal primer-probe sets without the loss of sensitivity. A comparison of the standard curves of viral RNA and IVT RNA showed that the IVT RNA quantified by ddPCR could be used as a standard for the absolute quantification of the qPCR assays. The results from the clinical samples showed that the ddPCR method could be used as a sensitive and quantitative diagnostic assay. Although the qPCR produced different Ct values using different primer-probes sets, the ddPCR results obtained using different primer-probe sets were compatible with each other. However, the narrow dynamic range and high cost of ddPCR are drawbacks for its use a clinical diagnostic method. Therefore, the deployment of qPCR assays with ddPCR-based reference materials could produce highly sensitive and quantitative results that are compatible between different laboratories.
Supplemental Materials
Acknowledgments
This work was supported by National Research Council of Science and Technology grant by the Ministry of Science and ICT (Grant No. CRC‐16‐01‐KRICT). This research was also supported by the “Establishment of measurement standards for Chemistry and Radiation”, grant number KRISS-2020-GP2020-0003, and “Development of Measurement Standards and Technology for Biomaterials and Medical Convergence”, grant number KRISS-2020-GP2020-0004 programs, funded by the Korea Research Institute of Standards and Science.
Conflict of Interest
The authors have no financial conflicts of interest to declare.
References
- 2015. WHO | Summary of probable SARS cases with onset of illness from 1 November 2002 to 31 July 2003. WHO.
- Fung TS, Liu DX. 2019. Human coronavirus: host-pathogen interaction.
Annu. Rev. Microbiol. 73 : 529-557. - Li W, Wong S-K, Li F, Kuhn JH, Huang I-C, Choe H,
et al . 2006. Animal origins of the severe acute respiratory syndrome coronavirus: insight from ACE2-S-protein interactions.J. Virol. 80 : 4211-4219. - Wu K, Peng G, Wilken M, Geraghty RJ, Li F. 2012. Mechanisms of host receptor adaptation by severe acute respiratory syndrome coronavirus.
J. Biol. Chem. 287 : 8904-8911. - Ge XY, Li JL, Yang X Lou, Chmura AA, Zhu G, Epstein JH,
et al . 2013. Isolation and characterization of a bat SARS-like coronavirus that uses the ACE2 receptor.Nature 503 : 535-538. - WHO. Middle East respiratory syndrome coronavirus (MERS-CoV). https://www.who.int/emergencies/mers-cov/en/.
- Ashour HM, Elkhatib WF, Rahman MM, Elshabrawy HA. 2020. Insights into the recent 2019 novel coronavirus (Sars-coV-2) in light of past human coronavirus outbreaks.
Pathogens 9 : 186. - World Health Organization (WHO). Novel Coronavirus ( 2019-nCoV ) Situation Report - 1 21 January 2020, WHO Bull. 1-7.
- Coronaviridae Study Group of the international committee on taxonomy of viruses 2020. 2020. The species severe acute respiratory syndrome-related coronavirus: classifying 2019-nCoV and naming it SARS-CoV-2.
Nat. Microbiol. 5 : 536-544. - da Costa VG, Moreli ML, Saivish MV. 2020. The emergence of SARS, MERS and novel SARS-2 coronaviruses in the 21st century.
Arch. Virol. 165 : 1517-1526. - Hindson BJ, Ness KD, Masquelier DA, Belgrader P, Heredia NJ, Makarewicz AJ,
et al . 2011. High-Throughput droplet digital PCR system for absolute quantitation of DNA copy number.Anal. Chem. 83 : 8604-8610. - Hindson CM, Chevillet JR, Briggs HA, Gallichotte EN, Ruf IK, Hindson BJ,
et al . 2013. Absolute quantification by droplet digital PCR versus analog real-time PCR.Nat. Methods 10 : 1003-1005. - Kinloch NN, Ritchie G, Brumme CJ, Dong W, Dong W, Lawson T,
et al . 2020. Suboptimal biological sampling as a probable cause of false-negative COVID-19 diagnostic test results.J. Infect. Dis. 222 : 899-902. - Alteri C, Cento V, Antonello M, Colagrossi L, Merli M, Ughi N,
et al . 2020. Detection and quantification of SARS-CoV-2 by droplet digital PCR in real-time PCR negative nasopharyngeal swabs from suspected COVID-19 patients.PLoS One 15 : e0236311. - Falzone L, Musso N, Gattuso G, Bongiorno D, Palermo CI, Scalia G,
et al . 2020. Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection.Int. J. Mol. Med. 46 : 957-964. - Suo T, Liu X, Feng J, Guo M, Hu W, Guo D,
et al . 2020. ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens.Emerg. Microbes Infect. 9 : 1259-1268. - Dang Y, Liu N, Tan C, Feng Y, Yuan X, Fan D. 2020. Comparison of qualitative and quantitative analyses of COVID-19 clinical samples.
Clin. Chim. Acta J. 510 : 613-616. - Dong L, Zhou J, Niu C, Wang Q, Pan Y, Sheng S,
et al . 2020. Highly accurate and sensitive diagnostic detection of SARS-CoV-2 by digital PCR.Talanta 224 : 121726. - Lu R, Wang J, Li M, Wang Y, Dong J, Cai W. 2020. SARS-CoV-2 detection using digital PCR for COVID-19 diagnosis, treatment monitoring and criteria for discharge.
medRxiv . 2020.03.24.20042689. - Gonzalez R, Curtis K, Bivins A, Bibby K, Weir MH, Yetka K,
et al . 2020. COVID-19 surveillance in Southeastern Virginia using wastewater-based epidemiology.Water Res. 186 : 116296. - Chang L, Yan Y, Zhao L, Hu G, Deng L, Su D,
et al . 2020. No evidence of SARS-CoV-2 RNA among blood donors: a multicenter study in Hubei, China.Transfusion 60 : 2038-2046. - Ahmed W, Bertsch PM, Angel N, Bibby K, Bivins A, Dierens L,
et al . 2020. Detection of SARS-CoV-2 RNA in commercial passenger aircraft and cruise ship wastewater: a surveillance tool for assessing the presence of COVID-19 infected travellers.J. Travel Med. 27 : 1-11. - Mostafa HH, Hardick J, Morehead E, Miller J, Gaydos CA, Manabe YC. 2020. Comparison of the analytical sensitivity of seven commonly used commercial SARS-CoV-2 automated molecular assays.
J. Clin. Virol. J. 130 : 104578. - Lv J, Yang J, Xue J, Zhu P, Liu L, Li S. 2020. Detection of SARS-CoV-2 RNA residue on object surfaces in nucleic acid testing laboratory using droplet digital PCR.
Sci. Total Environ. 742 : 140370. - Whale AS, Jones GM, Pavšič J, Dreo T, Redshaw N, Akyurek S,
et al . 2018. Assessment of digital PCR as a primary reference measurement procedure to support advances in precision medicine.Clin. Chem. 64 : 1296-1307. - Bhat S, Emslie KR. 2016. Digital polymerase chain reaction for characterisation of DNA reference materials.
Biomol. Detect. Quantif. 10 : 47-49. - Kwon HJ, Jeong JS, Bae YK, Choi K, Yang I. 2019. Stable isotope labeled DNA: a new strategy for the quantification of total dna using liquid chromatography-mass spectrometry.
Anal. Chem. 91 : 3936-3943. - Yoo HB, Park SR, Dong L, Wang J, Sui Z, Pavsic J,
et al . 2016. International comparison of enumeration-based quantification of DNA copy-concentration using flow cytometric counting and digital polymerase chain reaction.Anal. Chem. 88 : 12169-12176. - Corbisier P, Vincent S, Schimmel H, Kortekaas A-M, Trapmann S, Burns M,
et al . 2012. CCQM-K86/P113.1: relative quantification of genomic DNA fragments extracted from a biological tissue.Metrologia 49 : 8002-8002. - Miotke L, Lau BT, Rumma RT, Ji HP. 2014. High sensitivity detection and quantitation of DNA copy number and single nucleotide variants with single color droplet digital PCR.
Anal. Chem. 86 : 2618-2624. - Pinheiro LB, Coleman VA, Hindson CM, Herrmann J, Hindson BJ, Bhat S,
et al . 2012. Evaluation of a droplet digital polymerase chain reaction format for DNA copy number quantification.Anal. Chem. 84 : 1003-1011. - Martinez-Hernandez F, Garcia-Heredia I, Gomez ML, Maestre-Carballa L, Martínez JM, Martinez-Garcia M. 2019. Droplet digital PCR for estimating absolute abundances of widespread pelagibacter viruses.
Front. Microbiol. 10 : 1226. - Americo JL, Earl PL, Moss B. 2017. Droplet digital PCR for rapid enumeration of viral genomes and particles from cells and animals infected with orthopoxviruses.
Virology 511 : 19-22. - Hayden RT, Gu Z, Ingersoll J, Abdul-Ali D, Shi L, Pounds S,
et al . 2013. Comparison of droplet digital PCR to real-time PCR for quantitative detection of cytomegalovirus.J. Clin. Microbiol. 51 : 540-546. - Fronhoffs S, Totzke G, Stier S, Wernert N, Rothe M, Brüning T,
et al . 2002. A method for the rapid construction of cRNA standard curves in quantitative real-time reverse transcription polymerase chain reaction.Mol. Cell. Probes 16 : 99-110. - Kim DW, Kim YJ, Park SH, Yun MR, Yang JS, Kang HJ,
et al . 2016. Variations in spike glycoprotein gene of MERS-coV, South Korea, 2015.Emerg. Infect. Dis. 22 : 100-104. - Jung Y, Park G-S, Moon JH, Ku KB, Beak S-H, Lee C-S,
et al . 2020. Comparative analysis of primer-probe sets for RT-qPCR of COVID-19 causative virus (SARS-CoV-2).ACS Infect. Dis. 6 : 2513-2523. - Corman VM, Eckerle I, Bleicker T, Zaki A, Landt O, Eschbach-Bludau M,
et al . 2012. Detection of a novel human coronavirus by realtime reverse-transcription polymerase chain reaction.Euro. Surveill. 17 : 20285. - Vijgen L, Keyaerts E, Moës E, Maes P, Duson G, Van Ranst M. 2005. Development of one-step, real-time, quantitative reverse transcriptase PCR assays for absolute quantitation of human coronaviruses OC43 and 229E.
J. Clin. Microbiol. 43 : 5452-5456. - BioRad 2015. Rare mutation detection best practices guidelines.
- Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW,
et al . 2020. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.Euro. Surveil. 25 : 2000045. - Wu F, Zhao S, Yu B, Chen YM, Wang W, Song ZG,
et al . 2020. A new coronavirus associated with human respiratory disease in China.Nature 579 : 265-269. - Zhou P, Yang X Lou, Wang XG, Hu B, Zhang L, Zhang W,
et al . 2020. A pneumonia outbreak associated with a new coronavirus of probable bat origin.Nature 579 : 270-273. - Walls AC, Park YJ, Tortorici MA, Wall A, McGuire AT, Veesler D. 2020. Structure, Function, and antigenicity of the SARS-CoV-2 Spike Glycoprotein.
Cell 181 : 281-292. - Wrapp D, Wang N, Corbett KS, Goldsmith JA, Hsieh CL, Abiona O,
et al . 2020. Cryo-EM structure of the 2019-nCoV spike in the prefusion conformation.Science 367 : 1260-1263. - Zhu N, Zhang D, Wang W, Li X, Yang B, Song J,
et al . 2020. A novel coronavirus from patients with pneumonia in China, 2019.N. Engl. J. Med. 382 : 727-733. - Winter AK, Hegde ST. 2020. The important role of serology for COVID-19 control.
Lancet Infect. Dis. 20 : 758-759. - Lu R, Wu X, Wan Z, Li Y, Jin X, Zhang C. 2020. A novel reverse transcription loop-mediated isothermal amplification method for rapid detection of SARS-CoV-2.
Int. J. Mol. Sci. 21 : 2826. - Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW,
et al . 2020. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.Euro Surveill. 25 : 2000045. - Chu DKW, Pan Y, Cheng SMS, Hui KPY, Krishnan P, Liu Y,
et al . 2020. Molecular diagnosis of a novel Coronavirus (2019-nCoV) causing an outbreak of pneumonia.Clin. Chem. 66 : 549-555. - Bustin, S.A. 2020. T-qPCR testing of SARS-CoV-2 : a primer.
Int. J. Mol. Sci 21 : 3004. - Li H, Bai R, Zhao Z, Tao L, Ma M, Ji Z,
et al . 2018. Application of droplet digital PCR to detect the pathogens of infectious diseases.Biosci. Rep. 38 : BSR1170. - Dobnik D, Kogovšek P, Jakomin T, Košir N, Žnidarič MT, Leskovec M,
et al . 2019. Accurate quantification and characterization of adeno-associated viral vectors.Front. Microbiol. 10 : 1570. - Abachin E, Convers S, Falque S, Esson R, Mallet L, Nougarede N. 2018. Comparison of reverse-transcriptase qPCR and droplet digital PCR for the quantification of dengue virus nucleic acid.
Biologicals 52 : 49-54. - Kiselinova M, Pasternak AO, De Spiegelaere W, Vogelaers D, Berkhout B, Vandekerckhove L. 2014. Comparison of droplet digital PCR and seminested real-time PCR for quantification of cell-associated HIV-1 RNA.
PLoS One 9 : e85999. - Yu F, Yan L, Wang N, Yang S, Wang L, Tang Y,
et al . 2020. Quantitative Detection and Viral Load Analysis of SARS-CoV-2 in Infected Patients.Clin. Infect. Dis. 71 : 793-798. - Kim D, Lee J-Y, Yang J-S, Kim JW, Kim VN, Chang H. 2020. The architecture of SARS-CoV-2 transcriptome.
Cell 181 : 914-921. - Adamski MG, Gumann P, Baird AE. 2014. A method for quantitative analysis of standard and high-throughput qPCR expression data based on input sample quantity.
PLoS One 9 : e103917. - Gutiérrez-Aguirre I, Rački N, Dreo T, Ravnikar M. 2015. Droplet digital PCR for absolute quantification of pathogens.
Methods Mol. Biol. 1302 : 331-347. - Coronavirus Standards Working Group. The Joint Initiative for Metrology in Biology. https://jimb.stanford.edu/covid-19-standards.
- ISO. COVID-19 response: freely available ISO standards. https://www.iso.org/covid19.
Related articles in JMB
Article
Research article
J. Microbiol. Biotechnol. 2021; 31(3): 358-367
Published online March 28, 2021 https://doi.org/10.4014/jmb.2009.09006
Copyright © The Korean Society for Microbiology and Biotechnology.
Comparison of Digital PCR and Quantitative PCR with Various SARS-CoV-2 Primer-Probe Sets
Changwoo Park1,2,3, Jina Lee1,4, Zohaib ul Hassan1,3,5, Keun Bon Ku3, Seong-Jun Kim3, Hong Gi Kim3, Edmond Changkyun Park3,5,6, Gun-Soo Park3,7, Daeui Park3,8, Seung-Hwa Baek3,8, Dongju Park1,3,9, Jihye Lee10, Sangeun Jeon10, Seungtaek Kim10, Chang-Seop Lee11,12, Hee Min Yoo1,5*, and Seil Kim1,3,5*
1Microbiological Analysis Team, Biometrology Group, Korea Research Institute of Standards and Science (KRISS), Daejeon 34113, Republic of Korea
2National Research Laboratory of Molecular Microbiology and Toxicology, Department of Agricultural Biotechnology, Seoul National University, Seoul 08826, Republic of Korea
3Center for Convergent Research of Emerging Virus Infection, Korea Research Institute of Chemical Technology, Daejeon 34114, Republic of Korea
4College of Pharmacy, Chungnam National University, Daejeon 34134, Republic of Korea
5Department of Bio-Analytical Science, University of Science and Technology (UST), Daejeon 34113, Republic of Korea
6Research Center for Bioconvergence Analysis, Korea Basic Science Institute, Cheongju 28119, Republic of Korea
7Research Group of Food Processing, Korea Food Research Institute, Wanju-gun, Jeollabuk-do 55365, Republic of Korea
8Department of Predictive Toxicology, Korea Institute of Toxicology, Daejeon 34114, Republic of Korea
9Department of Biological Science, Chungnam National University College of Bioscience and Biotechnology, Daejeon 34134, Republic of Korea
10Zoonotic Virus Laboratory, Institut Pasteur Korea, Seongnam-si, Gyeonggi-do 13488, Republic of Korea
11Department of Internal Medicine, Jeonbuk National University Medical School, Jeonju 54986, Republic of Korea
12Biomedical Research Institute of Jeonbuk National University Hospital, Jeonju 54907, Republic of Korea
Correspondence to:Hee Min Yoo, hmy@kriss.re.kr
Seil Kim, stapler@kriss.re.kr
Abstract
The World Health Organization (WHO) has declared the coronavirus disease 2019 (COVID-19) as an international health emergency. Current diagnostic tests are based on the reverse transcription-quantitative polymerase chain reaction (RT-qPCR) method, which is the gold standard test that involves the amplification of viral RNA. However, the RT-qPCR assay has limitations in terms of sensitivity and quantification. In this study, we tested both qPCR and droplet digital PCR (ddPCR) to detect low amounts of viral RNA. The cycle threshold (CT) of the viral RNA by RT-PCR significantly varied according to the sequences of the primer and probe sets with in vitro transcript (IVT) RNA or viral RNA as templates, whereas the copy number of the viral RNA by ddPCR was effectively quantified with IVT RNA, cultured viral RNA, and RNA from clinical samples. Furthermore, the clinical samples were assayed via both methods, and the sensitivity of the ddPCR was determined to be equal to or more than that of the RT-qPCR. However, the ddPCR assay is more suitable for determining the copy number of reference materials. These findings suggest that the qPCR assay with the ddPCR defined reference materials could be used as a highly sensitive and compatible diagnostic method for viral RNA detection.
Keywords: COVID-19, SARS-CoV-2, reverse transcription-quantitative polymerase chain reaction (RT-qPCR), droplet digital PCR (ddPCR), nucleocapsid protein gene, envelope protein gene
Introduction
Coronaviruses have caused several notable respiratory disease outbreaks in humans, such as severe acute respiratory syndrome (SARS). The SARS coronavirus was the first fatal pathogenic human coronavirus to be identified and caused 8,096 infections and 774 deaths in 26 countries in 2003 [1]. Prior to the emergence of SARS-CoV, only two HCoVs (HCoV-229E and HCoV-OC43) were known, both of which caused mild respiratory symptoms [2]. Bats were the natural hosts of SARS-CoV, and civet cats were regarded as the intermediate host [3]. Extensive studies on bat coronaviruses showed that such coronaviruses could be potential human pathogens [4, 5]. MERS-CoV was first reported in 2012 and caused 2,494 infections and 858 deaths in 27 countries as of November 2019 (WHO, Middle East respiratory syndrome coronavirus (MERS-CoV)). MERS-CoV also originated from bat coronaviruses, and camels were the intermediate host [7].
According to the World Health Organization (WHO), the WHO China Country Office was informed of cases of pneumonia of unknown etiology in Wuhan City, Hubei Province, on December 31st, 2019 [8]. The novel coronavirus, currently termed SARS-CoV-2, was officially announced as the causative agent by the International Committee on Taxonomy of Viruses (ICTV) [9]. SARS-CoV-2, or severe acute respiratory syndrome coronavirus 2, has resulted in over fifty million laboratory-confirmed cases and over a million deaths as of November 2020 [10]. A viral genome sequence was released for immediate public health support via the community online resource virological.org on January 10th (Wuhan-Hu-1, GenBank accession number MN908947) [3], followed by four other genomes that were deposited on January 12th in the viral sequence database curated by the Global Initiative on Sharing All Influenza Data (GISAID). The genome sequences suggest the virus is closely related to the members of a viral species termed severe acute respiratory syndrome (SARS)-related CoV, a species defined by the agent of the 2002/03 outbreak of SARS in humans [4, 5]. The species also comprises of a large number of viruses mostly detected in Rhinolophid bats in Asia and Europe. For the past several decades, quantitative polymerase chain reactions (qPCR) have become the gold standard for quantifying relative gene expressions. On the other hand, the recently developed digital polymerase chain reaction (dPCR) enables a highly sensitive measurement and absolute quantitation of nucleic acids [11, 12]. Droplet dPCR (ddPCR) can provide absolute quantification of target DNA or RNA molecules without external reference materials. The PCR reaction mixture with target nucleic acid molecules are portioned into thousands of nanoliter-sized water-in-oil emulsion droplets. These droplets are amplified with PCR, and each individual droplet is analyzed. The droplets with a template are counted as PCR-positive droplets, and the droplets without template are PCR-negative droplets. The absolute quantification of target nucleic acids can be done based on the number of positive and negative droplets. ddPCR has an advantage for the clinical diagnosis of SARS-CoV-2 to reduce the false negative results [13], which could be highly sensitive compared to real time PCR (RT-PCR) [14-17]. Moreover, there are some research groups that have analyzed many samples from COVID-19 confirmed patients by droplet digital PCR (ddPCR) and RT-PCR based on two target genes (ORF1ab and N). The overall concept looks similar to analyzing a large number of COVID-19 patient samples using digital PCR [18, 19]. However, none of these groups applied the ddPCR reference value using the reference material in the patient sample to measure an accurate quantification and to achieve diagnostic reliability [13,20-24]. DNA or RNA quantities should be metrologically traceable to a reference [25, 26]. As of this writing, multiple nucleic acid quantitation methods have been developed, such as chemical analysis methods based on isotope-dilution mass spectrometry (IDMS), capillary electrophoresis (CE), and enumeration-based flow cytometric method (FCM) counting [27, 28]. These methods can be accurately calibrated with solutions of nucleic acids. In addition, an international comparison study was performed between national metrology institutes (NMI) using the ddPCR method [29]. Recently, the droplet digital PCR (ddPCR) method has emerged as a powerful analytical technique for clinical utility [30, 31]. For example, ddPCR can be used for the rapid enumeration of viral genomes and particles [32-34].
In the present case of SARS-CoV-2, virus isolates or samples from infected patients have yet to be made available to the international public health community. In this study, we concluded that the optimized conditions are required to increase the precision of ddPCR to develop reference materials with matrix conditions.
Materials and Methods
Construction of RNA Standards
Full-length N and E genes of SARS-CoV-2 (GeneBank: MN908947) were synthesized and cloned in the pET21a plasmid. The plasmids were amplified with the T7 promoter primer (5’-TAATACGACTCACTATAGGG-3’) and T7 terminator primer (5’-GCTAGTTATTGCTCAGCGG-3’). The amplicons were spectrophotometrically quantified at 260 nm. A total of 200 ng of the PCR product was used for the in vitro transcription (MEGAscript T7 Transcription kit; Thermo Fisher Scientific, USA), which was performed at 37°C overnight in a 20 μl reaction mixture containing 2 μl of reaction buffer, 2 μl of each nucleoside triphosphate, and 2 μl of enzyme mix. The template DNAs were removed by digestion with 2 U of Turbo DNase I for 15 min at 37°C. The RNAs were precipitated by adding 6 μl of 3 M sodium acetate and 150 μl of 98% ethanol, which was followed by a subsequent incubation at -20°C for 30 min. After 15 min of centrifugation at 17,000 rpm, the supernatant was removed and 200 μl of 70% ethanol was added. After another 10-min centrifugation at 17,000 rpm, the supernatant was removed, and the pellet was dissolved in 20 μl of RNase-free H2O (Takara Bio Inc., Japan). Quantitation of the RNAs was performed spectrophotometrically at 260 nm. The measurement of the RNA concentrations was performed in duplicate, and the concentration was then converted to the molecule number [35].
Cell and Virus RNA Extraction
The human cell lines were obtained from the relevant culture collections Vero E6 cells (ATCC CRL-1586) and MRC-5 cells (ATCC CCL-171) were obtained from the American Type Culture Collection (ATCC), and Huh-7 cells (JCRB 0403) were obtained from the Japan Cell Research Bank (JCRB). The cells were cultured at 37°C in a 5% CO2 incubator unless described otherwise.
The clinical isolate SARS-CoV-2 (NCCP 43326) was obtained from the National Culture Collection for Pathogens (NCCP). The strains were introduced into the Vero E6 cells (ATCC CRL-1586) and maintained in Dulbecco’s modified Eagle medium (DMEM) supplemented with 2% (v/v) fetal bovine serum (FBS) and 1%antibiotic-antimycotic solution (Thermo Fisher Scientific). The SARS coronavirus HKU-39849 (GenBank:AY278491.2, provided by Dr. Malik Peiris of the University of Hong Kong) was inoculated into the Vero E6 cells. Cytopathic effects were observed two days after inoculation and the viral titer was determined via a plaque assay. The patient-derived isolate MERS-CoV strain KNIH/002_05_2015 [36] was obtained from the KNIH and inoculated into cultured Huh-7 cells. The Huh-7 cells were maintained in Dulbecco’s modified Eagle’s medium (DMEM, HyClone, USA) supplemented with 10% fetal bovine serum (FBS, HyClone, USA) at 37°C in a 5%CO2 incubator. HCoV OC43 (ATCC VR1588) was obtained from the ATCC and inoculated into the MRC-5 cells. The culture media of the MRC-5 cells were Gibco Minimum Essential Media (MEM; Thermo Fisher Scientific) supplemented with 10% heat-inactivated Fetal Bovine Serum (FBS), 1% Sodium Pyruvate, 1% penicillin, and 1%non-essential amino acids. The inoculated cells were cultured with virus growth medium (MEM with 2% heat-inactivated FBS, 1% penicillin, and 1% non-essential amino acids) at 33°C. The supernatant of the culture medium was stored at -80°C until RNA extraction. The viral RNAs of SARS-CoV-2, MERS-CoV and HCoV-OC43 were extracted using the QIAamp Viral RNA Mini Kit (Qiagen, Germany) according to the manufacturer’s instructions. The viral RNA of SARS-CoV was extracted from the culture medium using the MagMAX-96 viral RNA isolation kit (Thermo Fisher Scientific) according to the manufacturer's instructions.
Clinical Samples and RNA Preparation
The clinical samples used in this study were collected from subjects according to registered protocols approved by the Institutional Review Board (IRB) of Jeonbuk National University Hospital with all patients having signed written informed consent forms (IRB registration number: CUH 2020-02-050-008). The clinical characteristics of the patients are shown in Table S1. Upper respiratory tract specimens (naso- and oropharyngeal swabs) from COVID-19 patients were suspended in a transport medium (eNAT; Copan, USA) and stored at -80°C until use. The RNA extraction from the clinical samples was performed using the QIAamp Viral RNA Mini Kit (Qiagen) according to the manufacturer’s instructions. The extracted viral RNA was used as a template for qPCR assays and cDNA synthesis. The qPCR assays with undiluted RNA were done using the Allplex SARS-CoV-2 Assay kit (Seegene, Korea) according to the manufacturer’s instruction. The Ct values from extracted RNA were shown in Table S1.
Primer and Probes
The primer-probe sets used in this study are listed in Table 1. For the detection of SARS-CoV-2, primer-probe sets of from the Centers for Disease Control and Prevention (CDC, USA), the University of Hong Kong (HKU, Hong Kong), the National Institute of Infectious Disease Department of Virology III (NIID, Japan), and the National Institute of Health (NIH, Thailand) were selected according to laboratory guidance from WHO [37]. The specific primer-probe sets for MERS-CoV and HCov-OC43 were used for the quantification of the viral RNA genomes [38, 39]. All primers and probes were synthesized by Neoprobe (Korea). All probes were labeled with the reporter molecule 6-carboxyfluorescein (FAM) at the 5'-end and with the quencher Black Hole Quencher 1 (BHQ-1) at the 3'-end.
-
Table 1 . Information of primers and probes used in the study..
Institute Gene Name Type* Sequence (5’ → 3’) Taxon Ref. CDC (USA) N 2019-nCoV_N1 F GACCCCAAAATCAGCGAAAT SARS-CoV-2 [37] R** TCTGGTTACTGCCAGTTGAATCTG P+ ACCCCGCATTACGTTTGGTGGACC 2019-nCoV_N2 F TTACAAACATTGGCCGCAAA SARS-CoV-2 R GCGCGACATTCCGAAGAA P ACAATTTGCCCCCAGCGCTTCAG NIID (Japan) NIID_2019-nCoV_N F AAATTTTGGGGACCAGGAAC SARS-CoV-2 R TGGCAGCTGTGTAGGTCAAC P ATGTCGCGCATTGGCATGGA NIH (Thailand) WH-NIC N-R F CGTTTGGTGGACCCTCAGAT SARS-CoV-2 R CCCCACTGCGTTCTCCATT P CAACTGGCAGTAACCA HKU (Hong Kong) HKU-N F TAATCAGACAAGGAACTGATTA Sarbeco R CGAAGGTGTGACTTCCATG P GCAAATTGTGCAATTTGCGG Charité (Germany) E E_Sarbeco F ACAGGTACGTTAATAGTTAATAGCGT Sarbeco [41] R ATATTGCAGCAGTACGCACACA P ACACTAGCCATCCTTACTGCGCTTCG University of Bonn Medical Centre (Germany) upE F GCAACGCGCGATTCAGTT MERS-CoV [38] R GCCTCTACACGGGACCCATA P CTCTTCACATAATCGCCCCGAGCTCG University of Leuven (Belgium) OC43 F ATGTTAGGCCGATAATTGAGGACTAT HCoV-OC43 [39] R AATGTAAAGATGGCCGCGTATT P CATACTCTGACGGTCACAAT *F: forward primer, R: reverse primer, P: Probe.
cDNA Synthesis of Viral RNA Genomes
The cDNAs of the clinical and cultured (SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43) viral RNA genomes were synthesized using the LunaScript RT SuperMix Kit (New England Biolabs, USA). The reaction mixture had a volume of 20μl and consisted of 4μl of 5X LunaScript RT Supermix, 1μl of RNA template, and 15 μl of distilled water. The reactions were carried out according to the manufacturer’s instructions. The synthesized cDNAs were serially diluted with distilled water and used as templates for qPCR and ddPCR without purification.
qPCR and Droplet Digital PCR Measurement
The templates were serially diluted to a few copy numbers/μl. To remove the bias from different reverse transcription steps, the same cDNAs from the viral RNA were used as templates for both qPCR and ddPCR. RT-qPCR analysis was performed on the StepOne and StepOnePlus Real-Time PCR system (Thermo Fisher Scientific). The reaction mixture had a volume of 25μl and consisted of 12.5μl of maxima Probe/ROX qPCR master mix (2X) (Thermo Fisher Scientific), 1μl of cDNA template, 1μl of 10μM forward primer, 1μl of 10μM reverse primer, 1μl of 5 μM probe labeled with FAM, and 8.5 μl of distilled water. The ddPCR analysis was performed using a QX200 system (BioRad Laboratories, USA). The reaction mixture had a volume of 20μl and consisted of 10μl of ddPCR supermix for probes (BioRad Laboratories), 1μl of cDNA template, 1 μl of 10μM forward primer, 1μl of 10μM reverse primer, 1μl of 5 μM probe labeled with FAM, and 6 μl of distilled water. All valid copy numbers were selected according to the manufacturer's instructions [40].
Data Analysis
The qPCR data were initially analyzed using the StepOne software (Thermo Fisher Scientific). Raw data (
Results
Quantification of Viral and IVT RNAs
In this study, the primer-probe sets of each E gene were used to quantify the viral genomic RNA of SARS-CoV-2 and related coronaviruses in triplicate (Fig. 1 and Table 1) [37-39,41]. SARS-CoV, MERS-CoV, and HCoV-OC43 were used a negative control for the assessment of SARS-CoV-2 specific primer-probe sets. The copy numbers of SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 viral RNA were measured using ddPCR, and the measured values of the viral genomic RNA were 8.04 × 105, 6.52 × 105, 9.37 × 104, and 9.57 × 104 copies/μl, respectively. The standard curves of qPCR and ddPCR are linear, indicating the amplification efficiency of both methods is consistent (Fig. 1). The quantified viral genomic RNAs of SARS-CoV, MERS-CoV, and HCoV-OC43 were used as negative templates for SARS-CoV-2 specific assays. To assess the quantification of the viral genomic RNA and partial synthetic viral RNA, in vitro transcript (IVT) N and E gene RNA were measured via qPCR and ddPCR in triplicate (Fig. 2). The copy numbers of N IVT RNA and E IVT RNA were measured as 1.8 × 109 and 3.47 × 109, respectively. The amplification efficiencies were calculated from the standard curves of SARS-CoV-2 viral RNA and IVT RNAs. In the qPCR reactions, each E gene primer-probe sets targeting the SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 showed Ct values of 28.64, 27.96, 26.96, and 27.69, respectively. In addition, the ddPCR reactions with each E gene primer-probe set targeting SARS-CoV-2, SARS-CoV, MERS-CoV, and HCoV-OC43 showed copy numbers of 2.68 × 106, 4.69 × 106, 7.85 × 105, and 2.66 × 105, respectively.
-
Figure 1.
Detection of four viruses based on E gene assays by qPCR and ddPCR. (A ) and (B ): The Ct value and copy number of SARS-CoV-2 and SARS-CoV using the E sarbecovirus assay. (C ) and (D ): The Ct value and copy number of MERS and OC43 virus using the upE and OC43_E assay, respectively. All template were performed in triplicate and points indicates average value.
-
Figure 2.
Detection of the IVT gene by qPCR and ddPCR. (A ): The Ct value and copy number of the IVT N gene using the 2019-nCoV_N2 assay. (B ): The Ct value and copy number of the IVT E gene using the E sarbecovirus assay. All template were performed in triplicate and points indicates average value.
Primer-Probes Validation
To assess the specificity and sensitivity of the five SARS-CoV-2 specific N gene targeting primer-probe sets (2019-nCoV_N1, 2019-nCoV_N2, NIID_2019-nCoV_N, HKU-N, WH-NIC N, the assays were done with SARS-CoV-2 RNA and other human pathogenic corona viral RNAs (SARS-CoV, MERS-CoV, and HCoV-OC43). The qPCR and ddPCR reactions using the N gene targeting primer-probes sets were positive in the SARS-CoV-2 genomic RNA in triplicate (Fig. 3 and Table 3). The reactions were negative with the other betacorona viral genomic RNAs such as SARS-CoV, MERS-CoV, and HCoV-OC43. The reaction with the E primer-probe sets was positive for each target viral RNA. The reactions using
-
Table 3 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays..
Dilution factor 2019-nCoV_N1 2019-nCoV_N2 NIID_2019-nCoV_N WH-NIC N HKU-N qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR 10-2 18.89 S* 18.87 S 18.14 S 21.40 S 20.79 S 10-3 22.16 21173 22.19 20340 21.63 15173 24.65 19993 24.31 18473 10-4 25.99 1907 26.06 1659 25.40 1177 28.52 1599 28.09 1553 10-5 30.66 135 30.60 132 29.93 60 33.11 105 32.68 105 10-6 34.12 11 34.57 8 34.91 7 36.71 5 36.71 12 R2 0.9968 0.9973 0.9943 0.9971 0.9978 *Saturated.
**All template were performed in triplicate and data in table indicates average value..
-
Figure 3.
Comparison of the N gene primer-probe sets for the qPCR and ddPCR methods. For the comparison of viral SARS-CoV-2, primer-probe sets from the Centers for Disease Control and Prevention (CDC, USA) (A andB ), the National Institute of Infectious Disease Department of Virology III (NIID, Japan) (C ), the University of Hong Kong (HKU, Hong Kong) (D ), and the National Institute of Health (NIH, Thailand) (E ) were used based on the results of Table 4. All template were performed in triplicate and points indicates average value.
IVT RNAs as a Standard for RT-qPCR
The R2 values of the viral RNA and IVT RNA (N genes) standard curve from the RT-qPCR assays were calculated as 0.9973 and 0.9987, respectively, whereas the R2 values of the viral RNA and IVT RNA (E genes) standard curve from the RT-qPCR assays were calculated as 0.9956 and 0.9993, respectively (Figs. 1 and 2 and Tables 2 and 3). These values indicate that the standard curves of both the viral and IVT RNAs are linear and accurate. The amplification efficiency of both the viral and IVT RNAs were calculated from the slopes of the standard curves. The efficiencies of both the viral and IVT RNAs according to the target genes were similar (Figs. 1 and 2). The similar amplification efficiencies show that the quantified IVT RNA can be used as a standard for RT-qPCR measurement of viral RNA.
-
Table 2 . Validation of the SARS-CoV-2 specific N gene targeting primer-probe sets used in the study..
Template SARS-CoV-2 (15 pg/μl) SARS-CoV (8.7 pg/μl) MERS-CoV (72 pg/μl) HCoV-OC43 (60 pg/μl) Primer-probes qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 2019-nCoV_N1 26.30 6123 Undetermined 0 Undetermined 0 Undetermined 0 2019-nCoV_N2 26.57 5340 Undetermined 2.2 Undetermined 4.3 39.09 4.8 NIID_2019-nCoV_N 26.89 3993 Undetermined 0 Undetermined 0 Undetermined 0 HKU-N 27.57 4863 27.24 6030 Undetermined 0 Undetermined 0.8 WH-NIC N 28.91 6153 Undetermined 0 Undetermined 0 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value..
Comparison of qPCR and ddPCR Assays for COVID-19 Clinical Samples
Both the qPCR and ddPCR assays were tested with five SARS-CoV-2 positive and three SARS-CoV-2 negative clinical samples. The clinical characteristics of the confirmed COVID-19 patients with symptoms are summarized in Table S1. The positive patients were determined in clinics with the qPCR assay. The results from the positive samples showed that both qPCR and ddPCR can detect up to approximately 10 copy/μl of SARS-CoV-2 genomic RNA (Table 5). The control assays without templates did not produce any signals from both the qPCR and ddPCR assays. However, the qPCR and ddPCR assays with three negative patient samples (N1, N3, and N5) showed a high Ct value or small amount of copy number in some reactions (Table 4). According to the qPCR and ddPCR results with templates of very low concentrations, the signals from assays with a template of very low concentration can be difficult to distinguish from background signals. These results indicated that the sensitivity of ddPCR can be equal to or more than that of the optimized qPCR.
-
Table 4 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays on negative patient sample..
Primer-Probes 2019-nCoV_N2 primer-probe set WH-NIC N primer-probe set Template qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) N1 (153ng/μl) Undetermined 2.8 Undetermined 0 N3 (101ng/μl) Undetermined 4.3 Undetermined 0 N5 (243ng/μl) Undetermined 0.5 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value..
-
Table 5 . Results of the Ct values by qPCR and copy number by ddPCR on patient sample..
Primer-probe 2019-nCoV_N2 primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 26.87 1051 35.03 7.2 35.15 2.4 Undetermined 0 35.05 7.53 10-2 30.52 82 37.03 1.33 Undetermined 0.73 Undetermined 0.53 Undetermined 1.2 10-3 33.81 5.93 Undetermined 0.67 Undetermined 0.6 Undetermined 0 Undetermined 1.07 Primer-probe WH-NIC N primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 28.77 836.00 36.11 5 35.88 6.80 Undetermined 1.40 35.86 7.67 10-2 32.25 107.33 38.54 3 38.52 1.93 Undetermined 0.00 37.02 1.87 10-3 35.95 10.40 Undetermined 2.40 38.27 2.07 38.44 1.27 Undetermined 0.47 *The initial concentration of RNA templates were 46.9, 21.1, 27.5, 17.2, and 15.9 ng/μl, respectively..
**All template were performed in triplicate and data in table indicates average value..
+Only one of the triplicate was positive (single droplet).
Discussion
The outbreak of COVID-19 is a pandemic threat caused by the emergence of SARS-CoV-2, a newly discovered human coronavirus [42, 43]. SARS-CoV-2 has four structural proteins, known as the S (spike), E (envelope), M (membrane), and N (nucleocapsid) proteins [44]. The SARS-CoV-2 S protein binds with angiotensin-converting enzyme 2 (ACE2) with high-affinity, which leads to clinical features of pneumonia in many patients [45, 46]. SARS-CoV-2 is highly contagious and has quickly spread across the world, which further emphasizes the essential role of diagnostics in the control of communicable diseases [47].
Currently, a reliable test (also the most widely used) for detecting acute infections of SARS-CoV-2 is the reverse transcription-quantitative polymerase chain reaction (RT-qPCR) method [48]. Multiple national laboratories have developed assays targeting conserved regions that could potentially detect pancoronaviruses [49, 50]. A previous study provided information such as the specific primer and probe sequence, thermal profile, and reagents for the RT-qPCR to optimize the reaction conditions for the detection of SARS-CoV-2 RNA [51]. ddPCR is a method that improves upon conventional PCR methods that can individually amplify and directly quantify pathogen RNA or DNA [52]. Due to the principle of ddPCR, the signal from each droplet with a template can be clearly distinguished from the noise from the droplet without a template. Several studies have compared RT-qPCR and ddPCR for the quantification of viral genomes [53-55].
In this study, SARS-CoV-2 RNA concentrations were measured using the qPCR and ddPCR methods. The qPCR results showed that the Ct value of a given template was greatly dependent on the individual primer-probe sets. However, the copy numbers determined by ddPCR were relatively stable and reliable in the quantification of viral RNA samples, regardless of the primer-probe sets. A previous study reported that qPCR and ddPCR results were consistent for 95 positive samples and that there was a high correlation between the Ct values of qPCR and the copy number values of ddPCR among patient samples [56]. Although the Ct values of each primer-probe set exhibited variance, the copy numbers from ddPCR showed that N genes are more abundant than E genes. These findings are also consistent with another previous study, indicating N genes are good targets for SARS-CoV-2 diagnostics [57].
The qPCR method requires the use of a standard curve to obtain quantitative measurements of an unknown sample [58]. However, in the case of ddPCR, the distribution of a target-specific amplification into partitions is calculated using the Poisson distribution, enabling the absolute quantification of the target gene based on the ratio of positive against all partitions at the end of the reaction without dependence on an external standard [59]. In addition to these advantages, our results revealed that ddPCR is a sensitive and accurate method for SARS-CoV-2 detection in low viral copies (Fig. 3). Further experiments should be performed even if patients are cured of COVID-19, and such tests should be conducted using ddPCR to lower the limit of detection (LOD) to be able to detect low amounts of viral RNA.
The Coronavirus Standards Working Group led by the Joint Initiative for Metrology in Biology (JIMB) is conducting international collaborative experiments to develop guidelines for the purpose of evaluating and establishing the availability of common and appropriate standards, diverse reference materials, validation tests, and reference measurement protocols that can enable reliable COVID-19 testing [60, 61]. Our results serve as a guide for the standardization of analytical methods and the development of reference materials that are required for accurate COVID-19 diagnosis based on both RT-qPCR and ddPCR. Although, ddPCR-based diagnostics can serve as a companion method to the current standard RT-qPCR to provide sensitive and accurate quantitative results for emergency testing that can be adopted for clinical use, there are some drawbacks. The narrow dynamic range and high capital and operational cost of ddPCR are the reasons preventing the deployment of ddPCR in clinical fields. Due to the nature of ddPCR, the precise quantification can be done with a decent number of positive droplets. ddPCR can be a precise diagnostics method, but there are some drawback. Therefore, the combination of both qPCR and ddPCR could be better methods for the SARS-CoV-2 diagnostics; qPCR assay with ddPCR defined reference materials could be precise and practical assays for pathogen detection.
Conclusion
Our study demonstrated that the Ct values of samples from qPCR assays were significantly variable depending on the sequences of primer and probe sets, whereas the copy numbers of the samples from the ddPCR assays were hardly affected by the sequences of the primer and probe sets. These results indicate that ddPCR can be performed with sub-optimal primer-probe sets without the loss of sensitivity. A comparison of the standard curves of viral RNA and IVT RNA showed that the IVT RNA quantified by ddPCR could be used as a standard for the absolute quantification of the qPCR assays. The results from the clinical samples showed that the ddPCR method could be used as a sensitive and quantitative diagnostic assay. Although the qPCR produced different Ct values using different primer-probes sets, the ddPCR results obtained using different primer-probe sets were compatible with each other. However, the narrow dynamic range and high cost of ddPCR are drawbacks for its use a clinical diagnostic method. Therefore, the deployment of qPCR assays with ddPCR-based reference materials could produce highly sensitive and quantitative results that are compatible between different laboratories.
Supplemental Materials
Acknowledgments
This work was supported by National Research Council of Science and Technology grant by the Ministry of Science and ICT (Grant No. CRC‐16‐01‐KRICT). This research was also supported by the “Establishment of measurement standards for Chemistry and Radiation”, grant number KRISS-2020-GP2020-0003, and “Development of Measurement Standards and Technology for Biomaterials and Medical Convergence”, grant number KRISS-2020-GP2020-0004 programs, funded by the Korea Research Institute of Standards and Science.
Conflict of Interest
The authors have no financial conflicts of interest to declare.
Fig 1.
Fig 2.
Fig 3.
-
Table 1 . Information of primers and probes used in the study..
Institute Gene Name Type* Sequence (5’ → 3’) Taxon Ref. CDC (USA) N 2019-nCoV_N1 F GACCCCAAAATCAGCGAAAT SARS-CoV-2 [37] R** TCTGGTTACTGCCAGTTGAATCTG P+ ACCCCGCATTACGTTTGGTGGACC 2019-nCoV_N2 F TTACAAACATTGGCCGCAAA SARS-CoV-2 R GCGCGACATTCCGAAGAA P ACAATTTGCCCCCAGCGCTTCAG NIID (Japan) NIID_2019-nCoV_N F AAATTTTGGGGACCAGGAAC SARS-CoV-2 R TGGCAGCTGTGTAGGTCAAC P ATGTCGCGCATTGGCATGGA NIH (Thailand) WH-NIC N-R F CGTTTGGTGGACCCTCAGAT SARS-CoV-2 R CCCCACTGCGTTCTCCATT P CAACTGGCAGTAACCA HKU (Hong Kong) HKU-N F TAATCAGACAAGGAACTGATTA Sarbeco R CGAAGGTGTGACTTCCATG P GCAAATTGTGCAATTTGCGG Charité (Germany) E E_Sarbeco F ACAGGTACGTTAATAGTTAATAGCGT Sarbeco [41] R ATATTGCAGCAGTACGCACACA P ACACTAGCCATCCTTACTGCGCTTCG University of Bonn Medical Centre (Germany) upE F GCAACGCGCGATTCAGTT MERS-CoV [38] R GCCTCTACACGGGACCCATA P CTCTTCACATAATCGCCCCGAGCTCG University of Leuven (Belgium) OC43 F ATGTTAGGCCGATAATTGAGGACTAT HCoV-OC43 [39] R AATGTAAAGATGGCCGCGTATT P CATACTCTGACGGTCACAAT *F: forward primer, R: reverse primer, P: Probe.
-
Table 2 . Validation of the SARS-CoV-2 specific N gene targeting primer-probe sets used in the study..
Template SARS-CoV-2 (15 pg/μl) SARS-CoV (8.7 pg/μl) MERS-CoV (72 pg/μl) HCoV-OC43 (60 pg/μl) Primer-probes qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 2019-nCoV_N1 26.30 6123 Undetermined 0 Undetermined 0 Undetermined 0 2019-nCoV_N2 26.57 5340 Undetermined 2.2 Undetermined 4.3 39.09 4.8 NIID_2019-nCoV_N 26.89 3993 Undetermined 0 Undetermined 0 Undetermined 0 HKU-N 27.57 4863 27.24 6030 Undetermined 0 Undetermined 0.8 WH-NIC N 28.91 6153 Undetermined 0 Undetermined 0 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value..
-
Table 3 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays..
Dilution factor 2019-nCoV_N1 2019-nCoV_N2 NIID_2019-nCoV_N WH-NIC N HKU-N qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR qPCR ddPCR 10-2 18.89 S* 18.87 S 18.14 S 21.40 S 20.79 S 10-3 22.16 21173 22.19 20340 21.63 15173 24.65 19993 24.31 18473 10-4 25.99 1907 26.06 1659 25.40 1177 28.52 1599 28.09 1553 10-5 30.66 135 30.60 132 29.93 60 33.11 105 32.68 105 10-6 34.12 11 34.57 8 34.91 7 36.71 5 36.71 12 R2 0.9968 0.9973 0.9943 0.9971 0.9978 *Saturated.
**All template were performed in triplicate and data in table indicates average value..
-
Table 4 . Comparison of the N-gene targeting SARS-CoV-2 qPCR and ddPCR assays on negative patient sample..
Primer-Probes 2019-nCoV_N2 primer-probe set WH-NIC N primer-probe set Template qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) N1 (153ng/μl) Undetermined 2.8 Undetermined 0 N3 (101ng/μl) Undetermined 4.3 Undetermined 0 N5 (243ng/μl) Undetermined 0.5 Undetermined 0.5 *All template were performed in triplicate and data in table indicates average value..
-
Table 5 . Results of the Ct values by qPCR and copy number by ddPCR on patient sample..
Primer-probe 2019-nCoV_N2 primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 26.87 1051 35.03 7.2 35.15 2.4 Undetermined 0 35.05 7.53 10-2 30.52 82 37.03 1.33 Undetermined 0.73 Undetermined 0.53 Undetermined 1.2 10-3 33.81 5.93 Undetermined 0.67 Undetermined 0.6 Undetermined 0 Undetermined 1.07 Primer-probe WH-NIC N primer-probe set Template P1 P2 P3 P4 P5 Dilution factor qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) qPCR (Ct) ddPCR (copy/μl) 10-1 28.77 836.00 36.11 5 35.88 6.80 Undetermined 1.40 35.86 7.67 10-2 32.25 107.33 38.54 3 38.52 1.93 Undetermined 0.00 37.02 1.87 10-3 35.95 10.40 Undetermined 2.40 38.27 2.07 38.44 1.27 Undetermined 0.47 *The initial concentration of RNA templates were 46.9, 21.1, 27.5, 17.2, and 15.9 ng/μl, respectively..
**All template were performed in triplicate and data in table indicates average value..
+Only one of the triplicate was positive (single droplet).
References
- 2015. WHO | Summary of probable SARS cases with onset of illness from 1 November 2002 to 31 July 2003. WHO.
- Fung TS, Liu DX. 2019. Human coronavirus: host-pathogen interaction.
Annu. Rev. Microbiol. 73 : 529-557. - Li W, Wong S-K, Li F, Kuhn JH, Huang I-C, Choe H,
et al . 2006. Animal origins of the severe acute respiratory syndrome coronavirus: insight from ACE2-S-protein interactions.J. Virol. 80 : 4211-4219. - Wu K, Peng G, Wilken M, Geraghty RJ, Li F. 2012. Mechanisms of host receptor adaptation by severe acute respiratory syndrome coronavirus.
J. Biol. Chem. 287 : 8904-8911. - Ge XY, Li JL, Yang X Lou, Chmura AA, Zhu G, Epstein JH,
et al . 2013. Isolation and characterization of a bat SARS-like coronavirus that uses the ACE2 receptor.Nature 503 : 535-538. - WHO. Middle East respiratory syndrome coronavirus (MERS-CoV). https://www.who.int/emergencies/mers-cov/en/.
- Ashour HM, Elkhatib WF, Rahman MM, Elshabrawy HA. 2020. Insights into the recent 2019 novel coronavirus (Sars-coV-2) in light of past human coronavirus outbreaks.
Pathogens 9 : 186. - World Health Organization (WHO). Novel Coronavirus ( 2019-nCoV ) Situation Report - 1 21 January 2020, WHO Bull. 1-7.
- Coronaviridae Study Group of the international committee on taxonomy of viruses 2020. 2020. The species severe acute respiratory syndrome-related coronavirus: classifying 2019-nCoV and naming it SARS-CoV-2.
Nat. Microbiol. 5 : 536-544. - da Costa VG, Moreli ML, Saivish MV. 2020. The emergence of SARS, MERS and novel SARS-2 coronaviruses in the 21st century.
Arch. Virol. 165 : 1517-1526. - Hindson BJ, Ness KD, Masquelier DA, Belgrader P, Heredia NJ, Makarewicz AJ,
et al . 2011. High-Throughput droplet digital PCR system for absolute quantitation of DNA copy number.Anal. Chem. 83 : 8604-8610. - Hindson CM, Chevillet JR, Briggs HA, Gallichotte EN, Ruf IK, Hindson BJ,
et al . 2013. Absolute quantification by droplet digital PCR versus analog real-time PCR.Nat. Methods 10 : 1003-1005. - Kinloch NN, Ritchie G, Brumme CJ, Dong W, Dong W, Lawson T,
et al . 2020. Suboptimal biological sampling as a probable cause of false-negative COVID-19 diagnostic test results.J. Infect. Dis. 222 : 899-902. - Alteri C, Cento V, Antonello M, Colagrossi L, Merli M, Ughi N,
et al . 2020. Detection and quantification of SARS-CoV-2 by droplet digital PCR in real-time PCR negative nasopharyngeal swabs from suspected COVID-19 patients.PLoS One 15 : e0236311. - Falzone L, Musso N, Gattuso G, Bongiorno D, Palermo CI, Scalia G,
et al . 2020. Sensitivity assessment of droplet digital PCR for SARS-CoV-2 detection.Int. J. Mol. Med. 46 : 957-964. - Suo T, Liu X, Feng J, Guo M, Hu W, Guo D,
et al . 2020. ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens.Emerg. Microbes Infect. 9 : 1259-1268. - Dang Y, Liu N, Tan C, Feng Y, Yuan X, Fan D. 2020. Comparison of qualitative and quantitative analyses of COVID-19 clinical samples.
Clin. Chim. Acta J. 510 : 613-616. - Dong L, Zhou J, Niu C, Wang Q, Pan Y, Sheng S,
et al . 2020. Highly accurate and sensitive diagnostic detection of SARS-CoV-2 by digital PCR.Talanta 224 : 121726. - Lu R, Wang J, Li M, Wang Y, Dong J, Cai W. 2020. SARS-CoV-2 detection using digital PCR for COVID-19 diagnosis, treatment monitoring and criteria for discharge.
medRxiv . 2020.03.24.20042689. - Gonzalez R, Curtis K, Bivins A, Bibby K, Weir MH, Yetka K,
et al . 2020. COVID-19 surveillance in Southeastern Virginia using wastewater-based epidemiology.Water Res. 186 : 116296. - Chang L, Yan Y, Zhao L, Hu G, Deng L, Su D,
et al . 2020. No evidence of SARS-CoV-2 RNA among blood donors: a multicenter study in Hubei, China.Transfusion 60 : 2038-2046. - Ahmed W, Bertsch PM, Angel N, Bibby K, Bivins A, Dierens L,
et al . 2020. Detection of SARS-CoV-2 RNA in commercial passenger aircraft and cruise ship wastewater: a surveillance tool for assessing the presence of COVID-19 infected travellers.J. Travel Med. 27 : 1-11. - Mostafa HH, Hardick J, Morehead E, Miller J, Gaydos CA, Manabe YC. 2020. Comparison of the analytical sensitivity of seven commonly used commercial SARS-CoV-2 automated molecular assays.
J. Clin. Virol. J. 130 : 104578. - Lv J, Yang J, Xue J, Zhu P, Liu L, Li S. 2020. Detection of SARS-CoV-2 RNA residue on object surfaces in nucleic acid testing laboratory using droplet digital PCR.
Sci. Total Environ. 742 : 140370. - Whale AS, Jones GM, Pavšič J, Dreo T, Redshaw N, Akyurek S,
et al . 2018. Assessment of digital PCR as a primary reference measurement procedure to support advances in precision medicine.Clin. Chem. 64 : 1296-1307. - Bhat S, Emslie KR. 2016. Digital polymerase chain reaction for characterisation of DNA reference materials.
Biomol. Detect. Quantif. 10 : 47-49. - Kwon HJ, Jeong JS, Bae YK, Choi K, Yang I. 2019. Stable isotope labeled DNA: a new strategy for the quantification of total dna using liquid chromatography-mass spectrometry.
Anal. Chem. 91 : 3936-3943. - Yoo HB, Park SR, Dong L, Wang J, Sui Z, Pavsic J,
et al . 2016. International comparison of enumeration-based quantification of DNA copy-concentration using flow cytometric counting and digital polymerase chain reaction.Anal. Chem. 88 : 12169-12176. - Corbisier P, Vincent S, Schimmel H, Kortekaas A-M, Trapmann S, Burns M,
et al . 2012. CCQM-K86/P113.1: relative quantification of genomic DNA fragments extracted from a biological tissue.Metrologia 49 : 8002-8002. - Miotke L, Lau BT, Rumma RT, Ji HP. 2014. High sensitivity detection and quantitation of DNA copy number and single nucleotide variants with single color droplet digital PCR.
Anal. Chem. 86 : 2618-2624. - Pinheiro LB, Coleman VA, Hindson CM, Herrmann J, Hindson BJ, Bhat S,
et al . 2012. Evaluation of a droplet digital polymerase chain reaction format for DNA copy number quantification.Anal. Chem. 84 : 1003-1011. - Martinez-Hernandez F, Garcia-Heredia I, Gomez ML, Maestre-Carballa L, Martínez JM, Martinez-Garcia M. 2019. Droplet digital PCR for estimating absolute abundances of widespread pelagibacter viruses.
Front. Microbiol. 10 : 1226. - Americo JL, Earl PL, Moss B. 2017. Droplet digital PCR for rapid enumeration of viral genomes and particles from cells and animals infected with orthopoxviruses.
Virology 511 : 19-22. - Hayden RT, Gu Z, Ingersoll J, Abdul-Ali D, Shi L, Pounds S,
et al . 2013. Comparison of droplet digital PCR to real-time PCR for quantitative detection of cytomegalovirus.J. Clin. Microbiol. 51 : 540-546. - Fronhoffs S, Totzke G, Stier S, Wernert N, Rothe M, Brüning T,
et al . 2002. A method for the rapid construction of cRNA standard curves in quantitative real-time reverse transcription polymerase chain reaction.Mol. Cell. Probes 16 : 99-110. - Kim DW, Kim YJ, Park SH, Yun MR, Yang JS, Kang HJ,
et al . 2016. Variations in spike glycoprotein gene of MERS-coV, South Korea, 2015.Emerg. Infect. Dis. 22 : 100-104. - Jung Y, Park G-S, Moon JH, Ku KB, Beak S-H, Lee C-S,
et al . 2020. Comparative analysis of primer-probe sets for RT-qPCR of COVID-19 causative virus (SARS-CoV-2).ACS Infect. Dis. 6 : 2513-2523. - Corman VM, Eckerle I, Bleicker T, Zaki A, Landt O, Eschbach-Bludau M,
et al . 2012. Detection of a novel human coronavirus by realtime reverse-transcription polymerase chain reaction.Euro. Surveill. 17 : 20285. - Vijgen L, Keyaerts E, Moës E, Maes P, Duson G, Van Ranst M. 2005. Development of one-step, real-time, quantitative reverse transcriptase PCR assays for absolute quantitation of human coronaviruses OC43 and 229E.
J. Clin. Microbiol. 43 : 5452-5456. - BioRad 2015. Rare mutation detection best practices guidelines.
- Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW,
et al . 2020. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.Euro. Surveil. 25 : 2000045. - Wu F, Zhao S, Yu B, Chen YM, Wang W, Song ZG,
et al . 2020. A new coronavirus associated with human respiratory disease in China.Nature 579 : 265-269. - Zhou P, Yang X Lou, Wang XG, Hu B, Zhang L, Zhang W,
et al . 2020. A pneumonia outbreak associated with a new coronavirus of probable bat origin.Nature 579 : 270-273. - Walls AC, Park YJ, Tortorici MA, Wall A, McGuire AT, Veesler D. 2020. Structure, Function, and antigenicity of the SARS-CoV-2 Spike Glycoprotein.
Cell 181 : 281-292. - Wrapp D, Wang N, Corbett KS, Goldsmith JA, Hsieh CL, Abiona O,
et al . 2020. Cryo-EM structure of the 2019-nCoV spike in the prefusion conformation.Science 367 : 1260-1263. - Zhu N, Zhang D, Wang W, Li X, Yang B, Song J,
et al . 2020. A novel coronavirus from patients with pneumonia in China, 2019.N. Engl. J. Med. 382 : 727-733. - Winter AK, Hegde ST. 2020. The important role of serology for COVID-19 control.
Lancet Infect. Dis. 20 : 758-759. - Lu R, Wu X, Wan Z, Li Y, Jin X, Zhang C. 2020. A novel reverse transcription loop-mediated isothermal amplification method for rapid detection of SARS-CoV-2.
Int. J. Mol. Sci. 21 : 2826. - Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DKW,
et al . 2020. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR.Euro Surveill. 25 : 2000045. - Chu DKW, Pan Y, Cheng SMS, Hui KPY, Krishnan P, Liu Y,
et al . 2020. Molecular diagnosis of a novel Coronavirus (2019-nCoV) causing an outbreak of pneumonia.Clin. Chem. 66 : 549-555. - Bustin, S.A. 2020. T-qPCR testing of SARS-CoV-2 : a primer.
Int. J. Mol. Sci 21 : 3004. - Li H, Bai R, Zhao Z, Tao L, Ma M, Ji Z,
et al . 2018. Application of droplet digital PCR to detect the pathogens of infectious diseases.Biosci. Rep. 38 : BSR1170. - Dobnik D, Kogovšek P, Jakomin T, Košir N, Žnidarič MT, Leskovec M,
et al . 2019. Accurate quantification and characterization of adeno-associated viral vectors.Front. Microbiol. 10 : 1570. - Abachin E, Convers S, Falque S, Esson R, Mallet L, Nougarede N. 2018. Comparison of reverse-transcriptase qPCR and droplet digital PCR for the quantification of dengue virus nucleic acid.
Biologicals 52 : 49-54. - Kiselinova M, Pasternak AO, De Spiegelaere W, Vogelaers D, Berkhout B, Vandekerckhove L. 2014. Comparison of droplet digital PCR and seminested real-time PCR for quantification of cell-associated HIV-1 RNA.
PLoS One 9 : e85999. - Yu F, Yan L, Wang N, Yang S, Wang L, Tang Y,
et al . 2020. Quantitative Detection and Viral Load Analysis of SARS-CoV-2 in Infected Patients.Clin. Infect. Dis. 71 : 793-798. - Kim D, Lee J-Y, Yang J-S, Kim JW, Kim VN, Chang H. 2020. The architecture of SARS-CoV-2 transcriptome.
Cell 181 : 914-921. - Adamski MG, Gumann P, Baird AE. 2014. A method for quantitative analysis of standard and high-throughput qPCR expression data based on input sample quantity.
PLoS One 9 : e103917. - Gutiérrez-Aguirre I, Rački N, Dreo T, Ravnikar M. 2015. Droplet digital PCR for absolute quantification of pathogens.
Methods Mol. Biol. 1302 : 331-347. - Coronavirus Standards Working Group. The Joint Initiative for Metrology in Biology. https://jimb.stanford.edu/covid-19-standards.
- ISO. COVID-19 response: freely available ISO standards. https://www.iso.org/covid19.